1、 Instrument Overview
Mainly used for measuring the impact toughness of non-metallic materials such as hard plastics, reinforced nylon, fiberglass, ceramics, cast stones, electrical insulation materials, and fiber reinforced thermosetting and thermoplastic materials. This testing machine is an ideal testing equipment for the chemical industry, scientific research institutions, colleges and universities, quality inspection departments, professional manufacturer laboratories, and other units.
2、 Instrument characteristics
This product has a wide measurement range, PLC touch screen control and display, automatic calculation of experimental results with USB storage data function, can be exported, edited and printed, with energy loss automatic correction function, clear and easy to operate, accurate and reliable measurement results.
3、 Compliant with standards
Standard number |
Standard name |
GB/T 1043.2-2018 |
《Determination of impact performance of plastic simply supported beams - Part 2: Instrumented impact test》 |
GB/T 18743-2002 |
《Test method for simply supported beam impact of thermoplastic pipes for fluid transport》 |
GB/T 3808-2018 |
《Inspection of pendulum impact testing machines》 |
JJG145-2007 |
《Pendulum impact testing machine》 |
GB/T 1843-2008 |
《Plastic cantilever beam impact test method》 |
ISO 180-2000 |
《Determination of Izod impact strength of plastics》 |
GB/T 2611-2007 |
《General technical requirements for testing machines》 |
4、 Sample type and size of simply supported beam (in mm)
Sample type |
Length(l) |
Width(b) |
thickness(d) |
1 |
80±2 |
10±0.5 |
4±0.2 |
2 |
50±1 |
6±0.2 |
4±0.2 |
3 |
120±2 |
15±0.5 |
10±0.5 |
4 |
125±2 |
13±0.5 |
13±0.5 |
5、 Cantilever beam specimen type size (in mm)
Sample model |
longL(mm) |
wideb(mm) |
thickh(mm) |
1 |
80±2 |
10.0±0.2 |
4 |
2 |
63.5±2 |
12.7±0.2 |
12.7±0.2 |
3 |
63.5±2 |
12.7±0.2 |
6.4±0.2 |
4 |
63.5±2 |
12.7±0.2 |
3.2±0.2 |
6、 Technical parameters
XJFD Composite impact testing machine |
|||
Parameters of simply supported beams |
Cantilever beam parameters |
||
Impact energy |
2J、4J、5J |
Impact energy |
2.75J、 5.5J |
Impact speed |
2.9m/s |
Impact speed |
3.5m/s |
Pendulum pre lift angle |
150° |
Pendulum pre lift angle |
150° |
Strike center distance |
230mm |
Strike center distance |
335mm |
Impact blade fillet radius |
(2±0.5)mm |
Blade fillet radius |
R=0.8mm±0.2mm |
source |
100VA 220VAC 50HZ |
||
External dimensions |
long420mm*wide250mm*650mm |
||
space required |
0.4m front and rear, 1.5m left and right, 1.5m upper |
7、 Instrument configuration
XJFD Composite impact testing machine |
|
main engine |
one |
Impact pendulum |
Two (2J, 2.75J) (orders below 1J need to be placed separately) |
Weights |
Three pairs (4J per pair, 5J per pair, 5.5J per pair) |
Composite jaws |
Two pairs |
Centering template |
one |
Inner hexagonal handle |
3mm、4mm、6mmOne each |
Power cord |
one |