Wansu testing for product quality escort
Wansu testing for product quality escort

了解详情

Product 当前位置:首页 > Product > XRL-II Methylene chloride impregnation testing machine
XRL-II Methylene chloride impregnation testing machine

1、 Overview
This instrument is used for conducting dichloromethane immersion tests on plastic pipes, fittings, sheets, etc. Widely used in production factories for pipes, fittings, and sheets, as well as in construction quality inspection and research departments.
2、 Compliant with standards
GB/T13526-2007: Test method for dichloromethane impregnation of rigid polyvinyl chloride (PVC-U) pipes.
ISO 9852-2007: Test method for dichloromethane impregnation of rigid polyvinyl chloride (PVC-U) pipes.
3、 Main components of the product
Composed of a constant temperature water bath, stainless steel container, sample fine hole filter, temperature control and timing system, etc.
4、 Technical indicators
1. Specification of stainless steel container: diameter 380mm, height 280mm
2. Constant temperature of dichloromethane: 20 ± 0.5 ℃
3. Instrument temperature control range: 10-50 ℃
4. Instrument temperature control accuracy: ± 0.2 ℃
5. Instrument display accuracy: 0.1 ℃
6. Timer range: 0-99 minutes
7. Timer accuracy: 1 second
5、 Sample standards
The sample should be cut to a length of at least 160mm.
6、 Sample method
1. The depth should submerge the sample by at least 30mm
2. The thickness of the water seal layer is about 20mm
3. The sample is kept at a constant temperature for 30 minutes, and after being taken out, it is dried for 15 minutes. The surface is severely damaged (such as sample peeling, cracking, scarring, etc.). The surface is damaged d. S
Slight change c. M
No change or slight change b. L
Severe surface damage (sample peeling, cracking, scarring, etc.)
The dichloromethane impregnation test is an important testing item for PVC-U pipes, used to characterize the degree of plasticization and uniformity of the pipes. The degree of plasticization of the pipes will directly affect their long-term static hydraulic performance and have a direct impact on their service life. Different standards also have different requirements for the testing conditions of the pipes.
7、 Reagents
Dichloromethane, analytical purity device: stainless steel container for 800 grit sandpaper constant temperature water tank timer
8、 Sample
Cut a pipe section at least 160mm long from the pipe, carefully turn the sample section with a right angle turning tool while minimizing material heating, and then gently polish with 800 grit sandpaper to make the section smooth and flat.
9、 Test steps
Pour dichloromethane into a container to a depth of at least 30mm to submerge the sample Slowly add distilled water to the container to form a 20mm thick water seal on dichloromethane.
Place the container in the thermostatic water tank, adjust the water bath temperature, when the temperature of dichloromethane reaches 20 ± 5 ℃, put the sample on the grid and immerse it in the container, or suspend the sample in dichloromethane, the immersion depth of the sample shall not be less than 30mm, and the polished surface of the sample shall be immersed in the solution. Soak the sample at a constant temperature for 20 minutes, remove it from the container, dry it for 15 minutes, visually inspect the inner and outer surfaces and cross-sections of the sample, measure the length of damage on the inner and outer surfaces and cross-sections with a tape measure and vernier caliper, and record all changes.
10、 Result representation
The categories of changes are evaluated using the following levels to assess the quality changes on the surface of the specimen: a N